HI 6600 Series Modular Sensor System User Guide
Page | 117
The screen shows the individual load sensors in dwell mode. Using the right+ left button,
move from Sensor to
Sensor. The pass/fail test requires 15-30 seconds to complete testing.
IT Dwell – Jbox1 Channel x (x/5)
RAW
Waversaver
Results OK
OK
Variation 0.45
0.00
mV/V
0.0219
Weight
-0.378605
RTZ
PASS
Integrated Technician Test for the HI 6600 over Communications
For Integrated Technician testing and data run over communications, there are two different tests the user can run:
The normal Integrated Technician test
The Reduced Voltage Integrated Technician test, which requires the new HI 6020IT summing card.
NOTE: The HI 6600 can use one HI 6020IT summing cards, with up to 4 load cells.
Once the test has been run, the unit will have the data collected during the test available. This data includes:
The number of load cells specified for each junction box.
Weight readings for each load cell and the junction box reference weight.
MV/V for each load cell and the junction box reference weight.
Raw variation for each load cell and the junction box reference weight.
WAVERSAVER® variation for each load cell and the junction box reference weight.
Raw variation results (pass/fail) for each load cell and junction box reference weight.
WAVERSAVER® variation results (pass/fail) for each load cell and junction box reference weight.
The Return to Zero (RTZ) results (pass/fail) for each load cell and reference weight as well as the combined
reading.
To run the test the user would need to know the parameter ID (a 32 bit hex value) for the two commands, Do IT Cmd and
Do IT Reduced (voltage) Cmd. These parameter IDs are:
16#00xx3197 Do IT Test Cmd
16#00xx3198 Do IT Reduced Cmd
The xx represents the Channel Number assigned to each Weight Processing Module. To run the test, the user would set
these parameters to a value of 1.
Once the IT Test has been run, the user can read each parameter ID they wish for the return value of that parameter. The
data will remain in the unit until the next time the IT test is run, which then would overwrite the previous data.
The parameter IDs are:
16#00xx3197 Do IT Test Cmd
16#00xx3197 Do IT Reduced Cmd