HI 6600 Series Modular Sensor System User Guide
Page | 77
6: WEIGHT SAMPLE CMD: Write a #6 to the command register to run the Weigh
Sample
command to set up the calibration of the Counts (if enabled).
o Status Error code 1 (motion)
o Status error code 2 (A/D error)
0x64 (100 decimal): CAL LOW CMD. Write a 0x64 hex to the command register to
perform the low
step of a traditional calibration. The status register will read 0 if this
command succeeds:
o Calibration_Fail 1
o Calibration_Fail_Motion 3
o Calibration_Fail_Adc_Error 4
o Code 2 indicates that the Calibration is in progress
0x65 (101decimal): CAL HIGH CMD. Write a 0x65 hex to the command register to
perform the high
step of a traditional calibration.
o
Calibration_Fail 1
o
Calibration_Fail_Motion 3
o
Calibration_Fail_Adc_Error 4
o HardcalFailCounts 8: not enough counts between Hard Cal High and Hard Cal Lo
o Code 2 indicates that the Calibration is in progress
0x66 (102 decimal): C2 CAL CMD. Write a 0x66 hex to the command register to
perform a C2
calibration.
o Calibration_Fail 1
o Calibration_Fail_Motion 3
o Calibration_Fail_Adc_Error 4
o Calibration_Fail_Noc2 5
o Calibration_Fail_C2capeq 6
o Calibration_Fail_C2clones 7
o Code 2 indicates that the Calibration is in progress
0x1000 (4096 decimal): WRITE INTEGER CMD. Set the value of an integer
parameter. Write 0x1000 in the command register 0, the parameter ID number in
registers 2, 3 and the desired value in registers 4, 5.
No Error Codes
0x1001 (4097 decimal): WRITE FLOAT CMD. Set the value of a floating point
parameter. Write 0x1001 in the command register 0, the parameter ID number in
registers 2, 3 and the desired float value in registers 4, 5.
Failure 1
TooHigh -1 : value is above legal limit
TooLow -2 : value is below legal limit
ErrorNotFound 0x8000 : parameter id not found
USB Memory Stick
The HI 6600 series provides an interface to an external USB memory stick that allows HI 6600 parameters to
be saved, modified, restored or copied to another instrument.
To begin, insert a USB stick into the Gateway Module, navigate to the Communications section of the Setup,
select USB, then select Save or Restore. The USB memory stick commands can be activated through the web