HI 6600 Series Modular Sensor System User Guide
Page | 119
16#00xx 49D7 RAW VARIATION RESULT CHAN7
16#00xx 49D9 RAW VARIATION RESULT JBOX REF2
16#00xx 49E0 WAVERSAVER® VARIATION RESULT CHAN0
16#00xx 49E1 WAVERSAVER® VARIATION RESULT CHAN1
16#00xx 49E2 WAVERSAVER® VARIATION RESULT CHAN2
16#00xx 49E3 WAVERSAVER® VARIATION RESULT CHAN3
16#00xx 49E8 WAVERSAVER® VARIATION RESULT JBOX REF1
16#00xx 49E4 WAVERSAVER® VARIATION RESULT CHAN4
16#00xx 49E5 WAVERSAVER® VARIATION RESULT CHAN5
16#00xx 49E6 WAVERSAVER® VARIATION RESULT CHAN6
16#00xx 49E7 WAVERSAVER® VARIATION RESULT CHAN7
16#00xx 49E9 WAVERSAVER® VARIATION RESULT JBOX REF2
16#00xx 49F0 RTZ CHAN0
16#00xx 49F1 RTZ CHAN1
16#00xx 49F2 RTZ CHAN2
16#00xx 49F3 RTZ CHAN3
16#00xx 49F8 RTZ JBOX REF1
16#00xx 49F4 RTZ CHAN4
16#00xx 49F5 RTZ CHAN5
16#00xx 49F6 RTZ CHAN6
16#00xx 49F7 RTZ CHAN7
16#00xx 49F9 RTZ JBOX REF2
16#00xx 498C RTZ COMBINED
Examples:
The user needs to run the IT test on channel 2. There are two methods of running the test on channel 2: 1) set the
parameter ID using the channel number and write a 1 to it; 2) set the base parameter number (channel number is 0) and
write to channel 2. We will show both examples here.
Example 1: Using the channel number in the parameter ID:
To run the test:
1. Enter the parameter ID to run the IT test for channel 2 into the Parameter ID register in the Common Header
section of the unit. This would set the 3
rd
register to a value of 16#0002 3197.
2. Enter the value of 1 (value to write) into the Parameter value register in the Common Header section of the unit.
This would set the 4
th
register to a value of 1.
3. Enter the Write command into the Command register in the Common Header section of the unit. This would set
the 1
st
register to a value of 16#1000.
4. The IT test will start running. Let this complete so all values will be available.
5. Set the Command Register in the Common Header section to a value of 0 (read command).
6. Enter the parameter ID 16#0002 4990 (parameter ID for the weight for 1
st
load cell on the 1
st
IT card) into
Parameter ID register in the Common Header section (3
rd
register).
7. Read the value returned in your input table, Parameter Value register (4
th
register) in the Common Header
section. NOTE: This will be a float data since this is a weight value being read.
8. Change the Parameter ID in the Output table to any other parameters you wish to read ensuring the channel
portion of the ID is set to the proper channel.
9. Read the value returned in the input table, parameter value register.
Example 2: Using the channel number in the Command Word:
To run the test:
1. Enter the parameter ID to run the IT test into the Parameter ID register in the Common Header section of the
unit. This would set the 3
rd
register to a value of 16#0000 3197.