1. select pump [Login]
2. select target for measurement [Login]
3. correct by 1ml [Login]
4. correct by 0.1ml [Login]
5. target measure [Login]
6. open lid [Login]
7. fill in solution
8. start calibration
9. back to service menu without calibration
1. calibration [Login]
2. user settings [Login]
3. device settings [Login]
4. network settings [Login]
5. back to system settings [Login]
10. indicate filling level [Login]
11. accept and save the calibration
12. discard calibration
8.6 Service menu
8.6.1 Calibration
Screen 1
Screen 2
1
5
4
3
2
1
2
3
4
5
6
7
8
9
11
12
10