Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland Operator_Manual_Rotolavit II_und_Rotolavit II-S_en_rev2.5
Seite 35
Down:
The number of revolutions per minute is adjustable
from 0 rpm to 3500 rpm. The time duration is
adjustable from 0 s to 20 s. The default value is
2000 rpm and 5 s.
Centrifugation run to centrifuge the remaining
droplets on the inner-side of the tubes to the bottom
of the tube.
Sedimentation:
The number of revolutions per minute is adjustable
from 0 rpm to 3500 rpm. The time duration is
adjustable from one second to two hours (0:00:01 to
2:00:00). The default value is 3500 rpm and 30
seconds (0:00:30)
The erythrocytes are sedimented by a selectable
rotation speed. The time only begins to count after
the set rotation speed is reached. After the time has
elapsed, a quick braking follows to prevent a
resuspension of the pellet.
The default value for the acceleration is 800 rpm/s.
The default value for the braking is 1000 rpm/s.
For the page change press onto the SPIN-icon
9.4.4 DOWN process
9.4.5 SPIN process
Page 1/2
Page 2/2
It is not possible to work with an endless spin process.
If it is necessary to work with a longer single spin process, it is possible to add a loop process for the
desired time-duration of max. 200 hours (with more processes to an absolute maximum of 3800 h)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich Rotolavit II and is the answer not in the manual?

Hettich Rotolavit II Specifications

General IconGeneral
BrandHettich
ModelRotolavit II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals