12.10 Maintenance schedule
Recommended minimum requirements. Regulations for your organization or physical conditions at your
organization may require maintenance items to be performed more frequently and or only by designated service
personnel.
Inspect the tubing and drain and clear obstructions if
necessary
Inspect the tubing connections and secure them if
necessary
Flush the system with deionized or distilled water
Clean and dry the interior after normal usage to prevent
corrosion and contamination
Flush the system with cleaning solution
Clean the fill ports on the rotor
Check the saline volume setting and calibrate it if
necessary. Frequency varies by length of service
Check the rotor speed and calibrate it if necessary
Inspect the rotor for wear, corrosion, and damage
Replace the rotor if these conditions exist
Inspect the tube holders for wear and damage
Replace tube holders if they are worn or damaged, or
after they have been in use for two years
Replace the supply and drain tubing
Replace the tube holder inserts for 10 mm x 75 mm
tubes
Replace the tube holders every two years
Replace the rotor, every four years (including the tube holders)