13.2 Software error codes
Program was successful, no error has
occurred
Program is still running, so far no error has
been detected (should never be visible in
history)
Motor could not be started (no speed could
be detected)
Motor is blocked
Motor cable connection
problem
Motor power supply problem
Motor could not accelerate within tolerance
(motor was too slow)
Wrong rotor type selected
Mechanical friction too big
Motor could not accelerate within tolerance
(motor was too fast)
Wrong rotor type selected
Motor could not hold the desired speed
(motor was too slow)
Wrong rotor type selected
Maximum of motor speed limit
too high (4000 rpm can maybe
not been hold)
Motor speed control does not
work as intended
Motor speed reading failure
Motor could not hold the desired speed
(motor was too fast)
Motor speed control does not
work as intended
Motor speed reading failure
Motor could not slow down within
tolerance
Wrong rotor type selected
Motor has signalized an error
Motor blocked
Motor over temperature
Motor power supply error
The 24 V of the motor supply is not
available
Lid is detected as open
Liquid injection system error
Pump was not able to pump the desired
amount of liquid
Pipe blocked
Pump not working
Flow sensor not working
Not enough liquid available or air is in the
pipe
Liquid container empty
Air in the pipe
Flow sensor problem
Open or check button was pressed, but
the lid could not be opened.
Lid mechanically blocked
Open or “CHECK” button was pressed, but
the lock could not be unlocked.
Motor was still rotating at the
time the command for
unlocking was received
Problem with the lock
Lid was opened without a request.
Emergency unlocking was
used