Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland Operator_Manual_Rotolavit II_und_Rotolavit II-S_en_rev2.5
Seite 33
9.3.8 wash 3 5ml 3x and anti (wash 3.5 ml, 3 x, plus antihuman globulin test)
The process values are:
FILL 3.5ml 1100rpm (fill 3.5ml by 1100rpm)
SPIN 20sec 3500rpm (acceleration 800rpm/s, braking 1000rpm/s)
DECANT 390rpm (decant by 390rpm)
AGIT 15x (15 agitations)
LOOP 3x (two more runs)
CHECK Pause (to add the antihuman globulin manually)
SPIN 20sec 3500rpm (acceleration 800rpm/s, braking 1000rpm/s)
9.3.9 wash white cells Tspot (wash leukocytes, Tspot)
The process values are:
FILL 2.5ml 900rpm (fill 2.5ml by 900rpm)
SPIN 7min 2260rpm (acceleration 800rpm/s, braking 1000rpm/s)
DECANT 370rpm (decant by 370rpm)
AGIT 100x (100 agitations)
LOOP 2x (only one more run)
9.3.10 cell recovery (only device type 1008-00S)
The process values are:
FILL 2.0ml 1100rpm (fill 2.0ml by 1100rpm)
SPIN 4min 2260rpm (acceleration 800rpm/s, braking 1000rpm/s)
DECANT 370rpm (decant by 370rpm)
AGIT 50x (50 agitations)
LOOP 2x (only one more run)
FILL 2.6ml 1100rpm (fill 2.6ml by 1100rpm)
9.2.11 immunophenotyping (only device type 1008-00S)
The process values are:
FILL 1.5ml 1100rpm (fill 1.5ml by 1100rpm)
SPIN 5min 1850rpm (acceleration 800rpm/s, braking 1000rpm/s)
DOWN 1100rpm
DECANT 370rpm (decant by 370rpm)
AGIT 15x (15 agitations)
LOOP 2x (only one more run)
FILL 0.5ml 1100rpm (fill 0,5ml by 1100rpm)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich Rotolavit II and is the answer not in the manual?

Hettich Rotolavit II Specifications

General IconGeneral
BrandHettich
ModelRotolavit II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals