Each command must be ended with ASCII code <CR>.
? Request for displayed value. Transmission format as described in chapter 11.1.
FNC x Choice of the measuring function
X= 012 3456 78
Func= —A -A B -B A+B A-B -A+B -A-B
FNC? Request for the measuring function
ID? Request for serial number of TE TT80, TE TT90 or TE UPC
MM / IN inch/metric conversion
MES? Request for the measurement-type, internal or external (INT, EXT)
PRE xxx Input of offset value for active function
xxx = true (x.xxxxx or -x.xxxxx in mm / x.xxxxxx or -x.xxxxxx en inch)
PRZ Zero-setting of offset value for each active measuring channel
RNG x Choice of the measuring range
X= 012 3456 78 9
Range µm ±5000 ±2000 ±500 ±200 ±50 ±20 ±5 ±2 ±0.5 Auto
Range in ±.200 ±.100 ±.020 ±.010 ±.002 ±.001 ±.0002 ±.0001 ±.00002 Auto
TT80
Range in/µin
±.100 ±.020 ±.010 ±5000 ±2000 ±1000 ±200 ±100 ±20 Auto
TT90
RNG? Request for the used range
RST Resets the instrument to its defaulted configuration (Reset)
STO0, STO1 Enables/Disables the measured value (HOLD function)
TOL xxx xxx Input of tolerance limits
xxx = true (x.xxxxx and -x.xxxxx in mm / x.xxxxxx and -x.xxxxxx in inch)
TOL? Request for tolerance limits
TO0, TO1 Enables/Disables the tolerance mode (in the TT90-UPC mode,
this command is ignored and the error message ERR23 displayed).
UNI? Request for actual unit : TT8 = mm / in – TT90 = µm / mil
VER? Request for the programme version
MEM x Selection of the measuring modes Min, Max, Max-Min or (Max+Min)/2
X = 012 34
Func. MEM= Normal Max Min Max-Min (Max+Min)/2
MEM? Request for the memory function
STM Sets the memories to actual value