Appendix
Page 73
heyer_LUVAR_UserManual_EN-3_CRev-1
Interference
emission
IEC 60601 test
level
Compliance level Electromagnetic envi-
ronment - guidelines
9 V/m
710, 745,
780MHz
PM: 217Hz
9 V/m
28 V/m
810, 870,
930MHz
PM: 18Hz
28 V/m
28 V/m
1720, 1845,
1970 MHz
PM: 217 Hz
28 V/m
28 V/m
2450 MHz
PM: 217 Hz
28 V/m
9 V/m
5240, 5500,
5785MHz
PM: 217 Hz
9 V/m
Fast transient
electrical
disturbances/
bursts accord-
ing to
IEC 61000-4-4
± 2 kV for pow-
er lines
± 1 kV for input
and output
lines
± 2 kV for power lines
± 1 kV for input and
output lines
The quality of the
supply voltage should
correspond to that of
a typical business or
hospital setting.

Table of Contents