199253
(6) -
EN
-
Accella™ Therapy Mattress Instructions for Use Page 19
Installing the Patient
Installing the Patient
Before Placing the Patient on the Device
Assess the various risks, including but not limited to the following (incomplete list
containing risks related to reasonably foreseeable misuse):
• risk of entrapment;
• potential falls from the bed;
• patient in state of confusion;
• patient's learning ability;
• persons lacking the mental capacity to recognize risky actions;
• unauthorized persons.
Install and use for the first time must be done in accordance with these
instructions.
Checking the Compatibility of Bed Frames and Mattresses
The bed / mattress / siderails combination (and in particular their respective
dimensions) must be examined in order to make sure that it meets the
requirements of the IEC 60601-2-52 standard and the ’’Hospital Bed Safety
Workgroup’’ guide, and that the resulting combination does not alter the
performance of the devices, their safety nor their usability.
Check that there are no overly large spaces on the sleep surface, especially by
the I-mmersion™ sensor, that could compromise the efficacy of the therapeutic
functions.