Accella™ Therapy Mattress Instructions for Use
Table of alerts
Table of the error codes on the GCI screen (combined mattress*)
Alarm silence
To silence alarms in case of mains power fault or malfunction, press the button .
The corresponding visual alarm remains on yellow.
The green indicator light near the symbol comes on.
The audible alarm is re-activated automatically after approximately 10 minutes.
Nevertheless, it is impossible to disable the bed exit alert.
The audible alarm can be stopped again for 10-minute periods until the problem
is solved. Refer to the Service Manual for more details.
Reason for sounding
Type of
alarm
Time before
sounding
Action to be taken by
the operator
At the end of the product
initialization phase
Audible (1
beep)
Immediately, at
the end of the
product
initialization
phase
Mattress ready for use
At the start of P-Max mode
Audible (1
beep)
Immediate Wait
At the end of P-Max mode
Audible (1
beep)
Immediately, at
the end of P-Max
mode
Repeat a P-Max, if
necessary
Patient exit
Audible
and visible
Immediate Attend to the patient
Microcontroller fault
Audible
and visible
Immediate
Switch off the alarm
and contact a Hill-Rom
technician
a
a. transfer the patient to a suitable support as quickly as possible
Error
Code
Fault description
1001 Pneumatic mattress pressure below 4.8 H20.
1002 Mattress pressure too high
1003 I-mmersion™ sensor fault
1004 Pressure fault on the MCB sensor
1005 I-mmersion™ sensor fault
1006 Pump(s) working continuously for more than 10 minutes
1007 Valve or deflation fault after P-Max stops
1008 Battery charger voltage out of range
1009 Low mattress battery voltage
1010 Fuse fault
1011 Fan fault
1012 Blower fault
1013 ALP mode fault
1014 P-Max pressure not reached after 22 minutes