Accella™ Therapy Mattress Instructions for Use
After 20 minutes, the device automatically returns to the initial therapy mode
to reduce the risks associated with non-therapy mode.
After activating the P-Max mode, it is possible to return to therapy mode
previously selected by selecting the P-Max button.
Transport mode
The Accella™ Therapy device is designed to stay inflated for
approximately 2 hours without power to ensure the support of the patient
during transport (See “Moving the Patient on the Bed in Transport Mode”
on page 33).
Locking
A lock function is available to prevent unexpected changes due to visitors.
Press to activate the lock.
All the LEDs on the remote control flash to confirm that the function is active.
If an attempt is made to activate other functions, the LEDs on the remote control
flash to indicate that they are locked.
NOTE:
Only the CPR function and the alarm silence cannot be locked out.
To deactivate the lock function, press again the button .
Patient exit function
This function is used to automatically detect when the patient leaves the
bed. This function is helpful for night staff or during busy periods and
increases patient safety.
Press the button to activate the patient monitor. The indicator light shines green.
If the patient leaves the bed, a light flashes next to the corresponding symbol and
a continuously alarm sounds.
Press again the button to switch off the patient monitor.
The patient exit function does not replace a suitable medical surveillance.
MCM™*
The microclimate management system is activated automatically when the
Accella™ Therapy device is switched on.
It can be switched off by pressing the Ventilation button on the remote control of
the MCM™* module.
When the MCM™ system is active, the green light comes on.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella P006783A and is the answer not in the manual?

Hillrom Accella P006783A Specifications

General IconGeneral
BrandHillrom
ModelAccella P006783A
CategoryMedical Equipment
LanguageEnglish

Related product manuals