Accella™ Therapy Mattress Instructions for Use Page 45
Disinfecting and
Servicing
Troubleshooting
The device is designed to adjust automatically, therefore troubleshooting is
limited to a few checks.
Always disconnect the device before troubleshooting.
In the event of a malfunction, press the silence alarm button to stop the audible
alarm. Check the following points:
• the connection to the mains is OK;
• the power cord is not damaged;
• the sensor cable is not damaged, nor disconnected;
• the device is not damaged (torn or punctured);
• the air filter is clean. Replace the air filter every 6 months.
Refer to the Service Manual for detailed troubleshooting instructions.
Replace the air filter (A),(B)
The air filter can be ordered separately. For more information, refer to the list
of spare parts of the mattress.
Replacing the Fuse on the Power Socket
1.Switch off the control
unit by pressing this
button.
2. Open the zip fastener.
3. Remove the power cord from the
connector.
4. Open the drawer of the
fuseholder with a small
screwdriver.
5. Check that the new fuse matches the characteristics on the label and meets
the IEC 60269-1 standard.
6. Replace the faulty fuse. The second fuse can be used to replace the faulty
fuse. Make sure to replace it in case of use.
7. Securely close the drawer of the fuseholder.
Others fuses are located on the MCB. Refer to Service Manual for
replacement and for verification.
If the problem is not solved, contact your local Hill-Rom representative (see back
of manual). Indicate the serial number of the device.