Accella™ Therapy Mattress Instructions for Use
For all other disinfectants:
Proceed with the final steps before reusing the cleaned and disinfected mattress
(see “Proceed with the final steps before reusing the cleaned and disinfected
device.” on page 43).
Cleaning and Disinfecting at Regular Intervals or in the Event of
High Risk of Contamination
Follow the same instructions as above (See “Cleaning and Disinfecting after the
Departure of a Patient or when Changing the Sheets” page 41), but with the
products listed below.
Using Recommended Products
• Only use compatible products and at the recommended concentrations (see
“List of compatible products” on page 40)
Dry Steam Cleaning the Device
Dry steam or superheated steam contains no more than 6% of water in
suspension and avoids the effects of condensation.
To avoid damage due to high pressure or an abnormal surface temperature,
take the following precautions:
• Use low pressure steam on the electric
parts.
• Do not use accessories such as high
pressure hoses (A) or soft non-metallic
brushes (B). It is advisable to only use
microfiber supports (C) on the cover, the
control pendant and the mains power cable
• Avoid water and steam getting into
connectors that are not in use.
• Do not brush and apply reduced pressure to labels and markings.
• Dry carefully and look for signs of water ingress.
• Test the device before reusing it.
Do not steam clean the inside parts but only the outside parts.
Do not wash the zip fasteners with steam.
Chlorine-based solutions can be used. Concentration must be
less than or equal to 1,000 ppm.
Ethanol-based solutions (alcohol) can be used. Concentration
must not exceed ¼ of ethanol for ¾ of water.