Accella™ Therapy Mattress Instructions for Use
Using cleaning and disinfecting protocols or products other than those
recommended by Hill-Rom can compromise the conformity of the device and
patient safety and render the warranty null and void.
Hill-Rom recommends that the device should be disinfected before scrapping, in
accordance with applicable local regulations.
Cleaning and disinfection are two separate procedures.
Products to be avoided
Never use cleaning agents, detergents, degreasing agents or industrial solvents
containing any of the following products in order to avoid damaging the mattress:
Do not use highly acidic detergents or disinfectants (pH<4).
Do not use highly basic detergents or disinfectants (pH>10).
Never use abrasive cleaning products or materials, such as scouring pads or
powders.
Recommended products
List of compatible products
Phenol
Hydrochloric,
nitric or sulfuric
acids
Dimethylformamide
Cresol Soda Tetrahydrofuran
Chemical class Active
Maximum
concentration
Chloride Sodium hypochlorite 0.1%
Alcohol Isopropyl alcohol 70%
Quaternary ammonium
n-Alkyl dimethyl benzyl
ammonium chloride
0.44%
Quaternary ammonium
chlorides
Didecyl dimethyl ammo-
nium chloride
0.2%
Peroxide Hydrogen peroxide 5%
Diamine
n-3-aminopropyl n-
dodecylpropane-1,3-
diamine
0.13%