Accella™ Therapy Mattress Instructions for Use
Servicing of the Device
Complying with Safety Instructions
Never modify this device without Hill-Rom's prior written consent.
Only facility-authorized personnel should perform maintenance.
Changes made by unauthorized personnel may result in damage to the
device and/or serious injury to staff or users.
Before maintenance or servicing works:
Ensure that the brakes are applied on the bed on which the mattress is
installed;
Lock out all electrical functions;
Disconnect the device;
Secure the mattress support platform and take the necessary steps to
prevent movement.
Do not work on the device when it is occupied
Refer to the Service Manual for assistance with the assembly, installation, use or
maintenance of the device. Contact your local Hill-Rom representative (see back
of manual) in the event of unforeseen events or behavior.
Performing Preventive Maintenance
A Service Manual and spare parts catalog are provided with the device. The
wiring diagrams, list of components, descriptions and calibration instructions can
be obtained on demand from Hill-Rom After-Sales.
The frequency of inspections must be adapted to the general condition of the
product and its use; for example, if the device is used by heavy patients. It is the
responsibility of the facility to implement a preventive maintenance program for
the device functions under its conditions of use.
The device and accessories should be inspected at least once a year to keep
them in good condition and working properly.
Every 3 years, it is recommended that Hill-Rom Customer Service or
a Hill-Rom-approved service provider inspects the device in order to maintain the
performance and safety of the apparatus over time.The battery must be replaced
during these inspections. The date of the next inspection must be recommended
according to the maintenance operations and observations.
In order to benefit from optimal and rapid service when calling Hill-Rom, provide
the serial number of the device for which you are calling (See “Locating the
Identification Labels of the Device” page 14).
The expected lifetime of the device in normal conditions of use, maintenance and
service is 5 years and 2 years for the cover.
Please refer to the service manual for more information on lifetime of wear
parts and accessories.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella P006783A and is the answer not in the manual?

Hillrom Accella P006783A Specifications

General IconGeneral
BrandHillrom
ModelAccella P006783A
CategoryMedical Equipment
LanguageEnglish

Related product manuals