Accella™ Therapy Mattress Instructions for Use Page 39
Disinfecting and
Servicing
Disinfecting and Servicing
Cleaning and Disinfecting the Device
Complying with Safety Instructions
• Check that the brakes are applied on the bed on which the mattress is
installed.
• Lock out all electrical functions.
• Disconnect the appliance and stow the power cable.
• Check that the connectors are securely connected to prevent water from
entering the mattress.
• Never clean with mattress by pouring water on it, or with a high-pressure jet.
• Never use water at a temperature of more than 70°C.
• Avoid excess water on the connectors.
• Refer to the recommendations of the cleaning and disinfecting product
manufacturer.
• Thoroughly dry the device before reusing it.
• Suitable individual protective equipment must be worn during the phases of
the cleaning operations (blouse, gloves, eye protection, etc.).
Failure to implement one or more of these recommendations may lead to
damage or deterioration, preventing use of the device and rendering the
warranty void.
Controlling Infection
Insufficient cleaning=Risk of infection (biological danger)!
All parts must be kept clean at all times in order to avoid the risk of
infection. All necessary precautions must be taken to eliminate all
visible soiling.
The following instructions are not designed to replace more appropriate
cleaning and disinfection protocols drawn up by the hygiene officer or by
other bodies for your hospital in the event of particular infectious
situations.
Complying with Hill-Rom Recommendations
The cleaning and disinfecting method described below applies specifically to this
device and its accessories and is designed to save time and to help combat
nosocomial infection more effectively.
Hill-Rom recommends that the device must be disinfected before first use.
When cleaning, always check the cover for cuts, tears, cracks or snags.
Never use a mattress with a damaged cover.