Page 20 199253
(6) -
EN
-
Accella™ Therapy Mattress Instructions for Use
Installing the Device
Standalone version
1. Unpack the control unit and the mattress.
Take care not to damage them when unpacking.
2. Check that all the components are present and correct, and that the power
cord is not damaged.
3. Place the rolled up mattress on the top of
the bed at head foot end and unroll it.
4. Check that the symbol on the cover
is at the foot of the bed.
5. Attach the mattress using the straps at
the head and foot ends.
6. Adjust the length of the straps to safely
secure the mattress.
7. Make sure that the device is correctly installed and securely attached, and in
particular that it is well centered on the sleep surface and pushed tightly
against the foot board to avoid the danger of entrapment.
8. Make sure that the attachments do not become entangled in the moving parts
of the bed chassis, such as actuators, CPR handles, etc. With articulated
chassis, make sure that the straps of the device are only attached to the
mobile head and foot sections and NOT to the main fixed chassis.