Accella™ Therapy Mattress Instructions for Use Page 23
Installing the Patient
8. Connect the plug to the bed connector (it clicks into place).
9. Install the hard surface of the thigh section.
10. Attach the mattress using the straps in the
center on the thigh section.
11. Unfold the mattress.
12. Adjust the length of the straps to securely
attach the mattress.
13. Make sure that the mattress is correctly
installed and securely attached, and in
particular that it is well centered on the
sleep surface and pushed firmly against
the foot board to avoid the danger of entrapment.
14. Connect the bed power cord to a wall outlet.
15. Press Mattress on the CGI
16. Press Start.
17. The mattress switches to initialization mode. A beep
sounds after 20 minutes, indicating that the mattress is
operational.
18. The status of the mattress switches to ON in the default
mode. The ALP and MCM™ modes are active.
19. Do not place the patient on the mattress during the initial inflation
phase before the ALP / MCM™ mode is activated. The operator must
check that all the zones of the mattress are fully inflated before
installation.
20. For small patients, center the sacrum on the I-mmersion™ Therapy zone
marked out on the mattress cover.
Placing a cotton sheet on the mattress will improve patient comfort and make
it easier to administer care.
It is advisable not to place any waterproof materials (e.g., drapes) on
mattresses with the MCM™ function, as doing so would reduce the
performance of the function.
Version combined with the Progressa™ bed
1. Unpack the control unit and the mattress.
Take care not to damage them when unpacking.
2. Check that all the components are present and correct, and that the power
cord is not damaged.