Accella™ Therapy Mattress Instructions for Use Page 33
Moving the Device
Moving the Device
Moving the Patient on the Bed in Transport Mode
1. Unplug the device from the wall socket.
Never pull on the power cable or you may damage it. A
damaged power cord is an electric shock hazard.
The device automatically switches to transport mode. The
pressure is distributed in the mattress until it is evenly inflated.
The mattress remains inflated.
A backup battery maintains for 2 hours functions such:
When the battery is low, the device must be connected to the mains for
at least 24 hours to fully charge, otherwise the functions associated
with the transport mode (range, availability CPR) are no longer
guaranteed.
The mains power fault alarm sounds and the yellow indicator light comes
on.
2. Deactivate the audible alarm by pressing (See “Alarm silence” page
30).
Audible alarm is re-activated automatically after approximately 10 minutes as
a reminder that the power cord must be reconnected.
3. Safely stow the power cord.
Make sure that the cord does not drag along the ground, do not
run over it when moving the bed chassis and take care not to trip
up over it. Where necessary, use the attachments provided with
the device.
4. Reconnect the device immediately following transport. It returns automatically
to the previous operating mode.
Standalone
version
Combined with
Accella™ bed
Combined with
Progressa™
bed
Emergency deflate
(CPR)
YES YES NO
Bed exit alarm YES NO NO
Low pressure alarm
YES YES
YES
(audible only)