142
────────────────────────────────────────────────────
5.6 Testing Using EXT I/O
────────────────────────────────────────────────────
Signal lines PNL2
──────
PNL1
─────
PNL0
──────
LD5
────
LD4
────
LD3
────
LD2
────
LD1
────
Settings HIGH LOW HIGH LOW HIGH LOW HIGH LOW
NOTE
Test values
AND result
Return to the
Initial screen
The first and third
arameters decision results
NOTE
3. When continuous testing is performed using panel numbers 6, 8 and 10, set
the following signals via the EXT I/O connector.
For details on each signals, panel saving pages, and panel numbers, see
Section 5.5.2, "Signal Lines for the EXT I/O Connector."
4. Input a trigger signal via EXT I/O connector to start test.
When a trigger signal (TRIG
--------------
) is input, the settings of LD1
----------
to LD5
----------
and PNL0
-------------
to PNL2
--------------
signals are read, and then the settings of panel numbers for
continuous testing are updated.
The testing will be performed in ascending order from low numbers of
specified panel. Each time test is completed, the test results for the panel
number are displayed.
The settings for the panel numbers on the screen are invalid because of
TRIG
---------------
signal input.
If the screen is switched to other pages after test completed, the displays of
the test results are disappeared.
The comparator decision result for the panel numbers for continuous testing,
which the comparator setting is ON, is output on the following signal line of
the EXT I/O connector.
AND
-------------
If all of the decision results are IN, then LO is output.
M-IN
-------------
If all of the decision results of the first parameters are IN, then LO
is output.
S-IN
-----------
If all of the decision results of the third parameters are IN, then LO
is output.
M-HI
-------------
,M-LO
---------------
,S-HI
-------------
,S-LO
-------------
Not used. HI is always output.