Chapter 4 Standard Measurement
___________________________________________________________________
A temperature indication "---" means that the temperature sensor
is not connected. The "---" indication may also appear if there
is a break in the test lead circuit.
Except for setting the auto power off mode and the moving
average function, do not press the
POWER
key in combination
with other keys. (For details of the moving average settings, see
Section 5.5, "Moving Average Function", and for details of the
auto power off settings, see Section 5.8, "Auto Power Off".)
If you do press a combination of the
POWER
key and other
keys, and an "
INSP
"or"
Adju
" indication appears, immediately
press the
POWER
key to power off and on again. Otherwise,
continuing with key operations may destroy the calibration data
for the unit, and correct measurement will no longer be possible.
If the leads are open-circuit, a spurious voltage indication may
sometimes be given. This is not a malfunction.