EasyManua.ls Logo

Hioki 3550 - Measurement Range Details; Resistance Measurement Range and Accuracy; Voltage Measurement Range and Accuracy

Hioki 3550
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
___________________________________________________________________
Chapter 3 Specifications
___________________________________________________________________
Range
Maximum
indication
Resolution
Measurement
current
Accuracy
30 m 30.00 m 10 µΩ 50 mA
Six months
1 year
0.8% rdg. 6 dgt.
1.2% rdg. 6 dgt.
300 m 300.0 m 100 µΩ 5mA
Six months
1 year
0.8% rdg. 6 dgt.
1.2% rdg. 6 dgt.
3 3.000 1m 500 µA
Six months
1 year
0.8% rdg. 6 dgt.
1.2% rdg. 6 dgt.
Range
Maximum
indication
Resolution Accuracy
3V 3.000 V 1mV
Six months
1 year
0.1% rdg. 6 dgt.
0.15% rdg. 6 dgt.
30 V 30.00 V 10 mV
Six months
1 year
0.1% rdg. 6 dgt.
0.15% rdg. 6 dgt.
Measurement range Resolution Accuracy
-10
o
Cto60
o
C 0.1
o
C
Six months
1 year
0.5% rdg. 10 dgt.
0.75% rdg. 15 dgt.
3.2 Measurement Ran
g
e
Conditions to guarantee accuracy:
Temperature :23
o
C 5
o
C
Humidity :80 %RH or less (no condensation)
Zero adjustment :After zero adjustment for each range
Warminguptime :Atleast10minutes
Period of guaranteed accuracy :1 year
(1) Resistance Measurement
Temperature coefficient: ( 0.01% rdg. 0.5 dgt.)/
o
C
Measurement current frequency: 1 kHz
30 Hz
(2) Voltage Measurement
Temperature coefficient: ( 0.005% rdg. 0.5dgt.)/
o
C
(3) Temperature Measurement

Other manuals for Hioki 3550