42
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
NOTE
5.7 Batter
Low Warnin
Shorting only the
SENSE
terminals will not display 0 V.
Always make sure that both the
SENSE
and
SOURCE
terminals
are shorted together.
If the leads are shorted but their ends are brought close to metal
parts, the measured value may fluctuate as a result of
electromagnetic induction. In this case, move the ends of the
leads away from the metal parts.
When the remaining battery capacity is low, the " "
indicator appears at the right of the display.
After printing out any held data values, replace the battery,
referring to Section 4.1, "Preparing for Measurement."
If the batteries are exhausted, you may be able to turn on the
power, but soon the "
" mark will appear and the power will go
off. (Momentary operation is possible because the battery partially
regains its former voltage after resting, but soon declines to the
exhausted state.) When the batteries wear out, replace them in
accordance with the specified procedure.