99
)XQFWLRQDO6SHFL¿FDWLRQV
(8) Auto Power Save function
Functionality The function automatically cuts off the power to the
instrument when it has been inactive for 10 minutes or
longer, and a constant current fault has been detected for
10 minutes or longer.
The function is disabled while the computer application is
sending and receiving data.
How to set Power-on option
Press the HOLD key while turning on the instrument to
enter the setup mode.
Select on or of
f and then press the ENTER key to apply the
settings.
(9) Buzzer function
Functionality The function sounds the buzzer based on the comparison
result of the comparator.
Default ON (The buzzer sounds if the comparison result is FAIL or
WARNING.)
How to set
Use the
key to turn it on and off (3 options).
OFF
ON (The buzzer sounds if the comparison result is PASS.)
ON (The buzzer sounds if the comparison result is FAIL or
WARNING.)