78
Accuracy Table
AC voltage
frequency
(
FREQ
U
)
Range
(Accuracy guarantee
range)
Resolution
Accuracy
Voltage frequency values are shown as “–
–”
when the voltage RMS value is less than 150
count. Voltage frequency values of less than
45.0 Hz are shown as “–
–.”
Maximum
display
999.9 Hz
(45.0 Hz to 999.9 Hz)
0.1 Hz
±0.3% rdg. ± 3 dgt.
999.9 Hz
(3) Single-phase power measurement
,
balanced 3-phase/4-wire power measurement
Auto range threshold: Range up: Current RMS value greater than 6000 count
Range down: Current RMS value less than 540 count
Effective
measuring range
Current RMS value
(
I
RMS
)
0.060 A to 600.0 A
Value must fall within the current measurement range’s guaranteed
accuracy range.
Voltage RMS value
(
U
RMS
)
80.0 V to 600.0 V
Frequency 50 Hz/60 Hz