Accuracy Table
Accuracy: dBV
Measurement
Range
[dBV]
20 to
below 45
[Hz]
45 to 65
[Hz]
Over 65
to 1 k
[Hz]
Over 1 k
to 10 k
[Hz]
Over 10 k
to 20 k
[Hz]
Over 20 k
to 100 k
[Hz]
-50 to
below -24
±0.8 dBV ±0.5 dBV ±0.5 dBV ±0.5 dBV ±0.6 dBV ±3.0 dBV
-24 to
below -4
±0.3 dBV ±0.2 dBV ±0.2 dBV ±0.2 dBV ±0.3 dBV ±1.0 dBV
-4 to below
15
±0.3 dBV ±0.2 dBV ±0.2 dBV ±0.2 dBV ±0.2 dBV ±0.5 dBV
15 to below
35
Not
specied
±0.2 dBV ±0.2 dBV ±0.2 dBV ±0.2 dBV ±0.5 dBV
35 to below
55
Not
specied
±0.2 dBV ±0.2 dBV ±0.2 dBV
Not
specied
Not
specied
55 to 60 Not
specied
±0.2 dBV ±0.2 dBV ±0.2 dBV
Not
specied
Not
specied
16
4-20 mA% conversion measurement
Range Conversion formula Accuracy
4-20 mA
350.00%
Measured value [mA] - 4 [mA]
16 [mA]
× 100
±0.1%rdg. ±20 dgt.
0-20 mA
300.00%
Measured value [mA]
20 [mA]
× 100
±0.1%rdg. ±20 dgt.
60.000 mA DC range xed