48
Measuring Conductances (DT4282)
3.10 Measuring Conductances
(DT4282)
The resistance is measured and the inverse number is displayed
(Unit: nS, nano-siemens). When the resistance is excessively large,
it is used.
Example: When the resistance is 50 M
Ω
, 1/50 M
Ω
= 20 nS.
(M = 10
6
, n = 10
-9
)
When the terminal is open, 0 nS appears.
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
3
4
Red
Black
1
2
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
The open terminal voltage is approx. 2.5 V or less. The
measurement current (DC) is approx. 96 nA. (p. 87)
To avoid damage to the measurement object, check the
specications before use.