87
Accuracy Table
8
Conductance (nS)
Range Accuracy Measurement current
600.00 nS ±1.5% rdg. ±10 dgt. 96 nA ±10%
• The accuracy is specied for humidity 60%RH or less.
• For more than 300 nS, ±20 dgt. is added.
• The accuracy is specied in 20.00 nS or more.
Open circuit voltage 2.5 V DC or less
9
Electrostatic capacity
Range Accuracy
*2
Measurement
current
Open circuit
voltage
1.000 nF ±1% rdg. ±20 dgt.
*1
32 μA ±10% 2.5 V DC or less
10.00 nF ±1% rdg. ±5 dgt.
*1
32 μA ±10% 2.5 V DC or less
100.0 nF ±1% rdg. ±5 dgt.
*1
32 μA ±10% 2.5 V DC or less
1.000 µF ±1% rdg. ±5 dgt. 32 μA ±10% 2.5 V DC or less
10.00 µF ±2% rdg. ±5 dgt.
*3
680 μA ±20% 3.1 V DC or less
100.0 µF ±2% rdg. ±5 dgt.
*3
680 μA ±20% 3.1 V DC or less
1.000 mF ±2% rdg. ±5 dgt.
*3
680 μA ±20% 2.1 V DC or less
10.00 mF ±2% rdg. ±5 dgt.
*3
680 μA ±20% 2.1 V DC or less
100.0 mF ±2% rdg. ±20 dgt.
*3
680 μA ±20% 2.1 V DC or less
*1: In the case of the 100 nF range or less, the accuracy is specied after the
REL function is activated.
*2: The accuracy is specied in 0.22 nF or more.
*3: Adds “Measurement accuracy × 0.3/°C” (except 23°C ±5°C)
• When the manual range is set, the accuracy is specied in 1% or more of
the range.
• The accuracy is specied for series resistance of 5
Ω
or less.
• Maximum count for each range: 1100 (excluding 100.0 mF)
1
2
3
4
5
6
7
Appx. Ind.