Accuracy Table
6
Diode test
Range Accuracy Measurement current
3.600 V ±0.1% rdg. ±5 dgt. 1.2 mA or less
Open circuit voltage 4.5 V DC or less
If the reading is lower than the threshold during the forward connection, a
buzzer sounds and red backlight comes on.
When the reading exceeds the threshold by 0.01 V or more during the
forward connection, the buzzer stops and the red backlight turns off.
Forward threshold
0.15 V/0.5 V (default)/1 V/1.5 V/2 V/2.5 V/3 V)
Threshold setting tolerance ±1% setting ±0.005 V
7
Resistance (
Ω
)
Range Accuracy Measurement current
60.000
Ω
±0.3% rdg. ±20 dgt.
*1
640 µA ±10%
600.00
Ω
±0.03% rdg. ±10 dgt.
*1
640 µA ±10%
6.0000k
Ω
±0.03% rdg. ±2 dgt.
*1
96 µA ±10%
60.000k
Ω
±0.03% rdg. ±2 dgt.
*1
9.3 µA ±10%
600.00k
Ω
±0.03% rdg. ±2 dgt. 0.96 µA ±10%
6.0000M
Ω
±0.15% rdg. ±4 dgt. 96 nA ±10%
60.00M
Ω
±1.5% rdg. ±10 dgt.
*2
96 nA ±10%
*5
600.0M
Ω
±3.0% rdg. ±20 dgt.
*2, *3
±8.0% rdg. ±20 dgt.
*2, *4
96 nA ±10%
*5
*1: Allow the input to short circuit and activate the relative value display
function (REL). The accuracy is then specied.
*2: The accuracy is specied for humidity 60%RH or less.
*3: The accuracy is specied for 200.00 M
Ω
or less.
*4: The accuracy is specied for more than 200.00 M
Ω
.
*5: The measurement current varies according to the resistance value of the
object under measurement.
Open circuit voltage 2.5 V DC or less
HIOKI DT4281A981-08