70
Using the Auto Power Save (APS)
4.11 Using the Auto Power Save (APS)
The auto power save function saves on power consumption. If the
instrument has not been operated for approx. 15 minutes, it enters
the sleep mode. (When the instrument is in the sleep mode, press
any key or turn the rotary switch to recover from the sleep mode.)
When the sleep mode continues for approx. 45 minutes, the power
turns off automatically.
Before shipping (as one of the default settings), the auto power
save function is set to enabled. ( lights up)
It is also possible to disable the auto power save function.
At 30 seconds before the instrument enters the sleep mode, the
APS blinks to indicate its status. To continuously use the instrument,
press any key or turn the rotary switch.
After use, set the rotary switch to OFF. The auto power save
function consumes a small amount of current.
Recovering from the power off state
Set the rotary switch to OFF and turn on the power again.
(When the current measurement is used, disconnect the measurement
cable and then turn on the power again.)
Disabling the auto power save function
TEMP C
TEMP FVERSION
RESET
0-20mA電源遮断断線(K熱電対)####
オートパワーセーブOFFブザーOFFメモリデータクリアメモリデータなし
Turn on the power while pressing
.
When is released, the measurement
display appears.
(APS goes off.)
This function is disabled until the power is turned
off.
HIOKI DT4281A981-08