Checking the Relative Value/Performing Zero Adjustment
4.7 Checking the Relative Value/
Performing Zero Adjustment
The relative value comparing to the standard value can be checked
(relative function).
It can also be used as the zero adjustment function.
Zero adjustment eliminates the inuences of the wiring resistance
(resistance measurement) and the wiring capacity (capacitor
measurement).
When the following measurement function is selected, this function
is disabled.
Peak measurement, Continuity, Diode, dBm/dBV, 4-20 mA
The measurement range cannot be changed when [REL] is lit. To
change the range, press REL for at least 1 second to reset the
REL function.
Checking the relative value (REL)
Example 1: DC voltage measurement
OVER
レンジオーバーCOPY: MEAS_
4-20
dBV
COPY: MEAS_
dBm
HOLDRANGE過負荷検出
OVER
最大入力範囲超え
ΔTRELMAXSLOW
When the standard value is measured,
press
for at least 1 second
(
lights up)
.
To cancel the state, press it for at least 1
second again. (
goes off.)
Example 2: Temperature measurement
OVER
レンジオーバーCOPY: MEAS_
4-20
dBV
COPY: MEAS_
dBm
HOLDRANGE過負荷検出
OVER
最大入力範囲超え
ΔTRELMAXSLOW
Deviation from the standard temperature
Standard temperature
When measuring temperature, [REF] lights up
and a standard temperature appears in the
sub display.