35
Inspection Before Use
2
Measure samples (such as battery, commercial power
supply, and resistor) of which values have already been
known, and check that the appropriate values appear.
Check method Action
Example:
Perform the AC voltage
measurement to measure the
commercial power supply, and then
check the display.
Red
Black
Normal:
An already-known value appears.
(In this example, the commercial
voltage level should appear.)
Abnormal:
The measured value does not
appear.
The malfunction may occur.
Stop the inspection and do not use
the instrument.
3
Check that the fuse is not broken.
Check method Action
1. Remove the fuse from the
instrument. (p. 106)
2. Reattach the battery cover.
3. In the resistance measurement,
check the resistance of the
fuse. (Resistance measurement
(p. 44))
Normal:
Fuse rating Resistance
630 mA Approx. 1.2
Ω
11 A 0.1
Ω
or less
Abnormal:
If the value above is not obtained (the
value higher than that is displayed),
replace the fuse. (p. 106)
1
2
3
4
5
6
7
Appx. Ind.