4.3 Starting and Stopping Recording
44
Depending on the selected recording interval, recording start is automatically delayed
until the next convenient clock time.
About Humidity Indications
To obtain correct humidity measurement values, the logger needs to wait
approximately 30 seconds for the measurement values to stabilize after
the screen display is turned on.
When using the logger for humidity monitoring or when calibrating the
logger, set the recording conditions so that the recording interval is 30
seconds or less, or turn off (disable) power saving.
Automatic Recording Start at Convenient Times
Recording Interval Recording Start Time
1sec 00 to 59 s (1-second interval)
2sec 00 to 58 s (2-seconds interval)
5sec 00 to 55 s (5-seconds interval)
10sec 00 to 50 s (10-seconds interval)
15sec 00 to 45 s (15-seconds interval)
20sec 00 to 40 s (20-seconds interval)
30sec 00 to 30 s (30-seconds interval)
1 min 00 min, 00 s to 59 min, 00 s (1-minute interval)
2 min 00 min, 00 s to 58 min, 00 s (2-minutes interval)
5 min 00 min, 00 s to 55 min, 00 s (5-minutes interval)
10 min 00 min, 00 s to 50 min, 00 s (10-minutes interval)
15 min 00 min, 00 s to 45 min, 00 s (15-minutes interval)
20 min 00 min, 00 s to 40 min, 00 s (20-minutes interval)
30 min 00 min, 00 s to 30 min, 00 s (30-minutes interval)
60 min 00 h, 00 min, 00 s to 23 h, 00 min, 00 s (1-hour interval)
At 12:01:00, you press the REC button.
The [REC] indicator in the display turns on,
but recording does not start yet.
At 12:10:00
Recording starts.
In the case of instantaneous value recording,
the number of data soon becomes 1.
In the case of statistical value recording, the
number of data becomes 1 at 12:20:00.
Example: When the button is pushed to start recording at 12:01:00, and
the recording interval is 10 minutes
12:01:00
Recording Start Operation
12:10:00
Actual Recording Start
If you try starting at 12:00 but are one minute late.
Actual recording starts at 12:10, which is considered the
next convenient time.