5.1 Scaling
63
3
5
Chapter 5 Processing Recorded Data
The following scaling calculation is applied to measured values.
Scaled Result = Raw data (measured value) × A + B × SI prefix (multiplier)
Scaled results are saved as a new item in the recording file.
1. Select the items, time span, and the following options.
* Set either one.
2. Confirm settings.
3. Click the [Exec
ute] button.
(The scaled results are saved.)
Note: Click the [Finish] button to close the [Scaling] dialog box.
5.1 Scaling
Item and range settings
Select the item to be scaled, and the
time span.
Clicking this tab
changes the setting
options. Make set-
tings on either tab.
(The settings are ap-
plied to the other
tab.)
A/B (slope/offset) values
1
2
3
1
1
Setting Options
Description
Specify by
example *
Enter two known conversion points (up to ten digits each).
Specify by A/B * Enter the scaling coefficients (A and B, up to ten digits each).
Scaled units
Select the [SI Prefix].
([p
]=1E-12, [n]=1E-9, []=1E-6, [m]=1E-3, blank =1E0, [k]=1E3, [M]=1E6,
[G]=1E9, [T]=1E12)
Enter a character string to identify the scaled units.
(Up to five characters, except \, /, :, *, ?, ", <, >, and |.)
Setting
confirmation
Confirm that scaling is performed properly. Enter any numerical value as raw
data, and click the [Calculate] button to display the scaled result.