9.4 Error Displays
95
3
9
Chapter 9 Maintenance and Service
The display appears as follows when an error occurs on the logger.
9.4 Error Displays
Logger Error Displays
Error Displays Meaning Remedies and References
Calibration data error:
A fault occurred with the internal
ca
libration data.
Inspection and repair is required.
Please contact your dealer or Hioki
r
epresentative.
See: "Requesting repairs" (p.91)
Microcomputer error:
A fault occurred in microcomputer
ROM/RAM.
Data recording error:
A fault occurred in recording data
or accessing settings.
Battery voltage is too low for nor-
mal logger operation.
Replace the battery.
See: "2.1 Installing (or Replacing)
the Battery" (p.17)
Measurement stopped during
recording because the battery is
depleted.
Replace the battery.
Resume recording after battery re-
placement.
See: "2.1 Installing (or Replacing)
the Battery" (p.17)
or
A measured value is out of range.
Out-of-range values cannot be dis-
played.
If the humidity display indicated O.F.,
d
ry the sensor and try again, as there
may be moisture on the sensor.
[OF] or [UF] is displayed when this
da
ta is imported by the LR5000 Utility
P
rogram
.
• The sensor plug is inserted
incorrectly.
• The sensor plug is not inserted
al
l the way in.
• The sensor is damaged.
• The logger is damaged.
Verify the correct plug orientation,
and insert it as far as possible.
If the values are not displayed despite
t
hese measures, the sensor and log-
ger need to be inspected and repaired.
Please contact your dealer or Hioki
r
epresentative.
See: "Requesting repairs" (p.91)
[ERROR]
is displayed when this
(faulty) data is imported by the
LR5000 Utility Program.
LR5000 Utility Program Error Displays
Error Displays Meaning Remedies and References
OF
A measured value is out of range.
Out-of-range values cannot be dis-
played.
UF