Model CP100NX CP90NX CP80NX
Dimensions
Width: 790 mm; Depth: 690 mm; Height: 880 mm
Depth with safety cover mounted: 890 mm
Height to the handle of the centrifuge door: 925 mm
Weight 390 kg
Power requirement**
Supply required
Single phase 50/60 Hz
208, 220 Vac+/-10 %, 20 A maximum (normally 8 A)
230, 240 Vac+/-10 %, 16 A maximum (normally 7 A)
Environmental conditions
Ambient temperature for operation: C to 40°C
Ambient Temperature for performance guarantee : 10°C to 30°C
Indoor use
Altitude up to 2000 m
Maximum relative humidity 80 % for temperatures up to 31 °C
decreasing linearly to 50 % relative humidity at 40 °C;
Pollution degree: 2
Overvoltage category
** The voltage to be used is the one that you specified when purchasing the centrifuge.
The CP100NX/90NX/80NX ultracentrifuges satisfy CE marking requirements. The CE marking is an
international symbol which shows that the product conforms to EC directives.
Standards concerning these directives are as follows:
Product safety (EN 61010-1 and EN 61010-2-020)
Electromagnetic compatibility (EN 61326-1)
1. Specifications
1-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hitachi CP100NX and is the answer not in the manual?

Hitachi CP100NX Specifications

General IconGeneral
BrandHitachi
ModelCP100NX
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals