1. Specifications
Model CP100NX CP90NX CP80NX
Maximum speed 100,000 rpm 90,000 rpm 80,000 rpm
Maximum RCF*
803,000 xg
700,000 xg
615,000 xg
Speed control accuracy
±2rpm(1,000rpm to maximum speed)
Acceleration/deceleration
control
10-stage variable acceleration control, 10-stage braked deceleration
control, plus coasting deceleration
Rotor temperature
control/display accuracy
±0.5 ℃(set temperature is from 0 ℃ to 40 ℃)
Set speed 1,000rpm to maximum speed in increments of 100 rpm
Set time
One minute to 999 hours and 59 minutes in increments of one minute,
“Hold” for continuous operation
Vacuum system
Oil rotary vacuum pump and oil diffusion pump combined
Ultimate vacuum: below 1.3 Pa (0.01 Torr) (This value reaches within 15
minutes after the rotor starts rotating.)
Noise level
51 dB (A scale)
(measured 1m in front of the instrument)
Maximum heat
dissipation into room
1 kW or less
Cooling method Thermo-module cooling(CFC/HCFC/HFC-free)
Screen display and
operation
Color touch-sensitive LCD (65,000 colors)
Interface USB X 2, LAN X1
Applicable rotors A rotor with an RLM adapter and a rotor with an optical disk
Rotor life management
Automatic management (A rotor with an RLM adapter)
(Like a rotor with an RLM adapter, a rotor with an optical disk will notify
when it reaches the end of its life as long as the rotor is registered and
correctly selected each time.
)
* RCF is an acronym for relative centrifugal force.
1-1