3-3 How to Use the Function Selection Area
This centrifuge provides programmed operation including step-mode operation, RCF display and setting
function, ω
2
T display and setting function and Real Time Control (RTC) that can run the centrifuge at a
preset date and time and a host of other functions. These features are displayed and controlled using the
Function Selection Area in the Run screen.
Name
Icon
Function
Program
You can save run parameters in memory for later use in repeated
operation. This feature also allows step-mode operation that can
combine a number of run parameters in a sequence of operations.
RCF
The centrifuge automatically computes and displays RCF values
from set speed, or speed from set RCF values.
ω
2
T
This function is used to indicate ω
2
T, (a value obtained by adding
run time to the angular velocity squared
), the centrifugal effect.
Operation can be stopped when the set ω
2
T value is obtained.
RTC
Sets a start time or end
time and runs the centrifuge at a desired
date and time.
The above features can be used in combination.
To combine [PROGRAM] and [RTC], first set [PROGRAM] and then [RTC]. Once
[RTC] is activated, you cannot change the run time and set [PROGRAM].
3-3 How to Use the Function Selection Area
Selection Area
3-19