Touchscreen operation Instrument operations and notes
10 The set time elapsed. (Time
out)
To stop before the set time
elapses, press the [STOP]
button.
When the set time elapses, the rotor decelerates and
stabilizes at the zonal speed.
The zonal operation chart uses a steady or blinking red light
to indicate the progress of the run.
When the speed stabilizes, a buzzer sounds
11 When the rotor stabilizes at
zonal speed, press the
[VACUUM] button.
The vacuum pump stops and the rotor chamber is vented to
return to normal atmospheric pressure.
The door lock is released.
12 Open the door, remove the
cap, install the seal assembly
and unload the samples.
CAUTION:
The cap is removed by hand from the rotor and the seal
assembly is installed when the rotor is turning. Perform
this operation with care and be sure to follow the
instructions in the instruction manual.
13 Press the [STOP] button.
Check that samples have been unloaded and press the
[YES] button.
The rotor decelerates and stops.
14 Remove the rotor.
Gently remove the rotor when it has stopped turning.
15 To stop zonal operation,
press the MENU tab and
press the [ZONAL setting]
button in the MENU screen.
16
Press the [NORMAL] button,
check that the button is now
surrounded by a yellow frame
before pressing the Run
The display again shows the Run screen.
The Zonal tab is hidden and the Function Selection Area
reappears.
3-4-3 Zonal Operation Procedures
●
asking whether it is
OK to decelerate and
stop.
●
screen appears.
(ZONAL Stop
3-69