LAB-X5000 Series User Guide
7.1.
Device type code LXDS2860
Rated voltage 100 V AC to 240 V AC
Rated current 1.5A
Rated frequency 47 Hz – 63 Hz
Fuse T 2A H 250 V
Average power consumption 60W
Operating temperature range +10 ˚C to +35 ˚C / +50 F to + 95 F
Storage temperature range -10 ˚C to +50 ˚C / +14 F to + 122 F
Recommended storage temperature Room temperature
Operating humidity range 20 % RH to 80 % RH
Maximum operating altitude 2000 m / 6500 ft
Embedded storage NAND 4 GB Flash Memory
Available storage space 2.3 GB
Maximum measurement count 100 000
Dimensions (W x L x H) 440 mm x 520 mm x 155 mm
Weight 20 kg
Pollution Degree PD2
Ingress Protection degree Not classified
Overvoltage category II
7.2.
Target material Palladium (Pd) or Titanium (Ti)
Window thickness 0.005 in
Anode voltage range 4 kV to 30 kV
Maximum anode current 0.75 mA
Maximum anode power 3.0 W
Nominal filament voltage 2.5 V
Maximum filament current 1.3 A
Minimum counts 24000
1/s
Ambient temperature range 0˚C to +40 ˚C / +32 F to + 104 F
Maximum package temperature 55 ˚C / +131 F