LAB-X5000 Series User Guide
Repeat for each element in the standard, until all elements included in the 5.4.6.10.
calibration have concentration values.
Press the Start button to analyze the standard. 5.4.6.11.
When analysis is complete, tap to add another standard. Repeat for each 5.4.6.12.
standard in the set.
Regression Analysis: When all standards have been input and analyzed, from the 5.4.7.
pull-down menu select Method, then tap Regressions. The computer will calculate
the regressions, then display “Calibration Complete”.
If there are not enough standards for the selected regression model, a warning will 5.4.7.1.
appear “More standards are needed for selected regression model”.
5.5.
Regressions
Select Method from the pull-down menu, then choose Regressions to perform a 5.5.1.
regression of the raw intensity data against the given concentration of the standards.
The error analysis for all elements in the analyzed standards will be calculated. The
minimum number of standards is 3. More are required when complex corrections are
applied.
There are four options for viewing the regression data. They can be toggled by 5.5.2.
tapping the
and buttons to cycle through the displays.
Graph 1: A graph of raw intensity (raw data) against concentration. 5.5.2.1.
Graph 2: A graph of given concentration against calculated concentration. 5.5.2.2.