INTRODUCTION
1
Imager Duo Operator’s Manual
1.13
Environmental
Operating Temperature Range
16°C to 32°C
Non-Operating Temperature Range
-28°C to 50°C
Operating Humidity Range
20 to 80% relative humidity, non-condensing
Non-Operating Humidity Range
15% to 95% relative humidity, non-condensing
Pollution Degree: II
, in accordance with IEC 61010-1.
Category II,
the ThinPrep Imaging System is for indoor use only in an office or
a clean laboratory environment.
Altitude:
0 meters (sea level) to 2000 meters.
Atmospheric Pressure:
1100 millibar to 500 millibar.
Power
Voltage:
100 - 240 Volts Alternating Current, no selection required
Mains supply voltage not to exceed ± 10% of the nominal voltage
Frequency:
50 to 60 Hz.
Power:
Imaging Station 4A maximum
Image Processor Controller Refer to the documents that come with the
equipment
Server Refer to the documents that come with the
equipment
Heat Generated:
Imaging Station approximately 1600 BTU/HR (470 W)
Image Processor ControllerRefer to the documents that come with the
equipment