EasyManua.ls Logo

Honeywell Chadwick-Helmuth 8500C - Figure C-1. Velocimeter and Magnetic Pickup Signal Phase Relationships

Honeywell Chadwick-Helmuth 8500C
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8500C/8500C+ System Maintenance Manual C-1
Appendix C. Chadwick-Helmuth Phase
Measurement Convention
When you use a signal generator to simulate signals to the 8500C/C+, it is important that you
understand how the 8500C/C+ measure phase. As shown in Figure C-1, a zero degree phase interval
measured on the 8500C/C+ occurs when the falling zero crossing of the vibration signal happens at the
same time as the magnetic pickup pulse. Therefore, on an H-P 3561A signal generator, a 180° phase
difference between two output signals would produce a 0° phase measurement on the 8500C/C+. A 0°
phase difference on the H-P would produce a 180° phase difference on the 8500C/C+. If you want to
adjust phase shift when using a signal generator, you must shift the vibration signal and not the
magnetic pickup signal.
Figure C-1. Velocimeter and Magnetic Pickup Signal Phase Relationships
FILE: SINWAVD.DRW
MAG P/U
SQUARE WAVE
SINEWAVE ( VIBRATION INPUT)
PHASE INTERVAL=0
MAG P/U
SQUARE WAVE
PHASE INTERVAL
SINEWAVE ( VIBRATION INPUT)
MAG P/U
SQUARE WAVE
PHASE INTERVAL
SINEWAVE ( VIBRATION INPUT)
MAG P/U
SQUARE WAVE
PHASE INTERVAL
SINEWAVE ( VIBRATION INPUT)
8500C PHASE INDICATION = 0 DEGREE ( 12:00 CLOCK)
(SIGNAL GENERATOR IS SET FOR 180 DEGREE PHASE SHIFT
BETWEEN TWO SIGNALS)
8500C PHASE INDICATION = 90 DEGREES ( 3:00 CLOCK)
8500C PHASE INDICATION = 180 DEGREES ( 6:00 CLOCK)
(SIGNAL GENERATOR IS SET FOR 0 DEGREE PHASE SHIFT
BETWEEN TWO SIGNALS)
8500C PHASE INDICATION = 270 DEGREES ( 9:00 CLOCK)

Table of Contents

Related product manuals