EasyManua.ls Logo

Honeywell Chadwick-Helmuth 8500C - Table 5-10. 8500 CC+ BalancerAnalyzer Calibration Tests

Honeywell Chadwick-Helmuth 8500C
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Guide
8500C/8500C+ System Maintenance Manual 5-39
Table 5-10. 8500C/C+ Balancer/Analyzer Calibration Tests
Problem Solution
5.5.1.3 Amplitude, Frequency, and Phase
Accuracy Tests
Vibration level does not meet specifications.
Frequency reading and/or phase angle do not
meet specifications.
1. Check 8500C/C+ harness assembly
(Figure 5-19) for shorted or open
connectors which can be caused by loose or
pinched wires, unplugged connectors, or
worn-through insulation.
2. Analog board is most likely at fault.
Replace analog board (Figure 5-17, A2).
1. Check 8500C/C+ harness assembly
(Figure 5-19) for shorted or open
connections which can be caused by loose
or pinched wires, unplugged connectors, or
worn-through insulation.
2. Digital board is most likely at fault. Replace
digital board (Figure 5-17, A1).
5.5.1.4 Noise and Spurious Signals Test
Noise above 0.005 IPS is present on display. 1. Turn EL backlight (Figure 2-1) all the way
down. Wait ten seconds and retest. Repeat
five times.
2. Check 8500C/C+ harness assembly
(Figure 5-19) for shorted or open
connections which can be caused by loose
or pinched wires, unplugged connectors, or
worn-through insulation.
3. Analog board is most likely at fault.
Replace analog board (Figure 5-17, A2).
5.5.1.5.1 500 Hz Anti-Aliasing
Signal aliasing above 0.005 IPS remains on
display.
1. Check 8500C/C+ harness assembly
(Figure 5-19) for shorted or open
connections which can be caused by loose
or pinched wires, unplugged connectors, or
worn-through insulation.
2. If noise is not at 500 Hz, it is not an aliased
signal; however, the noise and spurious
signals test (Section 5.5.1.4) should be
performed.

Table of Contents

Related product manuals