Security Definitions
E-Mon® Series - Global User Manual
54-000-01-ENG01-0121
Page | 73
9. Security Definitions
The security section enlists certain factors to protect the integrity of the E-Meter during
installation and operation. The users at all time shall be aware about data security and
integrity of the meter parts such as firmware, hardware and front panel display. Kindly adhere
to the highest standards during meter operation and installation.
9.1. Calibration and Tests
9.1.1. Calibration
The E-Mon meters are intended for factory installation only. The terminals have not
been evaluated for field-wiring applications.
The meters have been evaluated for use in Overvoltage Category III and intended to be
installed on the load side of suitable branch circuit protection.
9.1.2. Testing
The accuracy of the measurement function of the components has been investigated
under ANSI C12.20.
These meters have been tested for operation between -40°C to 70°C ambient
temperature. Additional testing is necessary if the meter is used outside the given
temperature range.
These meters have not been tested for tamper resistance for general operation.
Provide tamper detection, tamper resistance, or take measures to seal the meters, it
shall be determined in the final application.
9.2. Hardware
9.2.1. Sealable Terminals
Figure 25. Sealings

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell E-Mon EM3S-V-P-R and is the answer not in the manual?

Honeywell E-Mon EM3S-V-P-R Specifications

General IconGeneral
BrandHoneywell
ModelE-Mon EM3S-V-P-R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals