61
GasAlertMaxXT II
User Manual
Troubleshooting
If a problem occurs, refer to the solutions provided in Table 15. If the problem persists, contact BW Technologies by Honeywell.
Figure 14. Troubleshooting
Problem Possible Cause Solution
The detector does not activate. Depleted battery Charge the battery. Refer to Charging the
Battery.
Damaged or defective detector Contact BW Technologies by Honeywell.
Detector automatically deactivates. Automatic deactivation due to depleted
battery
Charge the battery. Refer to Charging the
Battery.
Lockout on Self-Test Error is enabled and
a sensor(s) has failed the startup self-test
Refer to Lockout on Self-Test Error
and
Replacing a Sensor or Sensor Filter
.
The detector enters alarm
immediately when activated.
Sensor needs to stabilize Used sensor: Wait 60 seconds
New sensor: Wait 5 minutes
Low battery alarm Charge the battery. Refer to Charging the
Battery.
Sensor(s) requires calibration Calibrate the sensor(s). Refer to
Calibration
.
Hazardous environment Leave the area immediately. Deactivate
and reactivate the detector in a safe area
that is free of hazardous gas.
The activation startup self-test fails. General fault Contact BW Technologies by Honeywell
.
Sensor error Replace the sensor. Refer to Replacing a
Sensor or Sensor Filter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell GasAlert Max XT II and is the answer not in the manual?

Honeywell GasAlert Max XT II Specifications

General IconGeneral
BrandHoneywell
ModelGasAlert Max XT II
CategorySecurity Sensors
LanguageEnglish

Related product manuals