GasAlertMaxXT II
User Manual
48
3. To ensure the readings are accurate, apply the verifica-
tion gas for the same amount of time as was applied to
the sensor when it was calibrated.
Example: H
2
S span time 2 minutes therefore, apply
verification gas for 2 minutes.
Calibrating Using the IR Link
If the Cal IR option is enabled, the sensors must be calibrated using
either the IR Link or MicroDock II station.
To calibrate using the IR Link, complete the following:
1. From the PC, open Fleet Manager II.
2. Click Administration.
3. From the Administration toolbar, click Login / Logout
to access the Enter Password dialog box. Enter Admin
and click OK.
4. From the Devices toolbar, click Detector Configuration.
5. Activate the detector.
Calibration can be performed during the startup tests
when CAL DUE NOW displays, or when the detector in
normal operation.
6. From Fleet Manager II, click the Calibrate button located
at the bottom of the window.
The Calibrate Device dialog box displays.
Figure 7. Calibrate Device Dialog Box

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell GasAlert Max XT II and is the answer not in the manual?

Honeywell GasAlert Max XT II Specifications

General IconGeneral
BrandHoneywell
ModelGasAlert Max XT II
CategorySecurity Sensors
LanguageEnglish

Related product manuals