Troubleshooting
If
you
cannot
restore
raw
files
If
you
cannot
restore
raw
files
If the
setup.
raw
and
data.
raw
files
seem
to be ignored when you copy
them to
the
logic analyzer:
D Verify
the
LAN
connection
between
your
computer
and
the
logic
analyzer. Refer
to
"If
you
cannot
connect
to
the logic analyzer" earlier in
this section.
D Change
the
file time-out. To change
the
time-out:
1 Go
to
the
logic analyzer System
External
1/0 menu.
2 Select
the
LAN
Settings field.
3 Set the file time out.
A value
of
about
1.5
seconds
should
be
sufficient. Set
the
time
out
to
the
minimum value
that
will
allow
network
operations
to
work
reliably.
If
you
cannot
telnet
to
the
command
parser
socket
If you get a "connection refused" message:
D Verify
that
you
included
the
telnet
port
number
(5025)
in
the
command.
If you get a "connection timed out" or "no response from host" message:
D Verify
the
LAN
connection
between
your
computer
and
the
logic
analyzer. Refer
to
"If
you
cannot
connect
to
the logic analyzer" earlier in
this section.
If you get a "connection refused" or "no response from host" message:
D
If
the
logic analyzer was
just
turned
on, make sure
that
you
wait 15
seconds after
the
Analyzer Configuration menu
is
displayed
before
attempting
the
connection.
D Make
sure
that
another
user
is
not
already accessing
the
file
system
as
the
control
user
or
connect.ed
to
the
command parser socket.
7-24